100/180 mm PAPERLESS GRAPHIC RECORDER: USER GUIDE
User Guide
Page 385
HA028910
Issue 13 Sep 16
PREVIOUS INSTRUMENTS
This section contains dc performance and resistance input data for instruments with status levels prior to
D6 (June 2007).
DC input ranges
Shunt
Externally mounted resistor modules
Additional error due to shunt
0.1% of input
Performance
See table
Low
High
Typical error
Maximum error
Worst case temperature
Range
Range
(Instrument at 20 deg C)
(Instrument at 20 deg C)
performance
-38 mV
38 mV
1.4
µ
V
0.035% input + 0.031% range 0.085% input + 0.052% range 80ppm of input per deg C
-150 mV
150 mV
5.5
µ
V
0.035% input + 0.028% range 0.084% input + 0.039% range 80ppm of input per deg C
-1 V
1 V
37
µ
V
0.035% input + 0.024% range 0.084% input + 0.029% range 80ppm of input per deg C
-20 V
20 V
720
µ
V
0.097% input + 0.027% range 0.448% input + 0.033% range 443ppm of input per deg C
Resolution
Note: The table above applies to recorders with status level prior to E7 (June 2007). For instru-
ments with a status level of E7 or above, see the
equivalent table
earlier in Annex A.
*WARNING
For portable case instruments only:
All I/O connections must be Low Voltage (maximum 33V ac RMS, 46.7V ac peak or 70Vdc) unless
the integrity of the instrument safety earth is maintained for as long as the I/O is connected to the
recorder.
Resistance inputs
Temperature scale
ITS90
Types, ranges and accuracies
See tables (values exclude infl uence of lead resistance)
Infl uence of lead resistance
Error:
Negligible
Mismatch:
1Ω/Ω
Maximum source current
250µA
Low
Range
0
Ω
0
Ω
0
Ω
High
Range
150
Ω
600
Ω
6k
Ω
Resolu-
tion
5m
Ω
22m
Ω
148m
Ω
Typical error
(Instrument at 20 deg. C)
0.027% input + 0.034% range
35ppm of input per deg C
35ppm of input per deg C
35ppm of input per deg C
0.027% input + 0.035% range
0.030% input + 0.028% range
Worst case temperature
performance
Maximum error
(Instrument at 20 deg. C)
0.042% input + 0.110% range
0.042% input + 0.053% range
0.045% input + 0.035% range
Note: The table above applies to recorders with status level prior to E7 (June 2007). For instru-
ments with a status level of E7 or above, see the
equivalent table
earlier in Annex A.
RTD type
Cu10
JPT100
Ni100
Ni120
Overall range (˚C)
Standard
0.02˚C
0.01˚C
0.01˚C
0.01˚C
0.01˚C
0.09˚C
0.01˚C
Pt100
Pt100A
Pt1000
-20 to + 400
-220 to + 630
-60 to + 250
-50 to + 170
-200 to + 850
-200 to + 600
-200 to + 850
General Electric Co.
JIS C1604:1989
DIN43760:1987
DIN43760:1987
IEC751
Eurotherm Recorders SA
IEC751
Max. linearisation error
Cu53
< 0.01˚C
-70 to + 200
RC21-4-1966