4.3.18 Demand Writes
DEMAND WRITE CONFIGURATION
Demand writes allow slave values to be written from the master by operator action (from the confi gura-
tion page), by job action, or (if the Screen Builder option is fi tted -
section 7
) by operation of a suitably
confi gured Operator key. Figure 4.3.18a shows a typical confi guration page. Modbus codes 6 and 16 can
be selected for this purpose - see
section 8.2.1
for code defi nitions.
Figure 4.3.18a shows a typical confi guration page for writing to a recorder. Figure 4.3.18b shows the Edit
Constant page. Figure 4.3.18c shows a confi guration page for the case where the destination is defi ned
as a register location.
WRITING TO A KNOWN DEVICE
Figure 4.3.18a shows the confi guration page for writing a value from the master unit to a ‘known’ slave
unit.
Figure 4.3.18a Typical Demand write confi guration page
Enable
Demand Type
Status
Master Comms
Apply
Discard
Write number
1) Write_1
Select the Write num-
ber to be configured
Select 'Master
Comms' or 'Off'
OK
Demand Write
Slave
1) Furn 1 Recorder
Select destination
slave
Parameter
Alarm setpoint 2
Select parameter
to be written-to
Point Type
Channel
Select destination
Point Number
1
Descriptor
Write 1
Source
Constant
Enter default value
Default
0
Allow Constant Edits
Disable Retries
Send On Power Up
On Error Write Default
Allows user to edit the constant prior
to write (if Source = 'constant')
If enabled, the default value is written, should
the selected source be in an 'Error' state.
If enabled, the write is attempted only once.
Write is triggered when the unit powers up.