4-38 Maintenance
Spray Manifold Disassembly Procedure
1
Press the
Power Off Control via the
membrane keypad.
2
Turn off the power at the facility power
safety disconnect for the Aquastorm
200
™
.
NOTE
It is important that power is
disconnected so the recirculating
pumps cannot be activated while
servicing the spray manifolds.
3
Once it is determined which of the spray
manifolds is to be removed for servicing,
open the front glass window, then remove
the splash sheild from inside and set it in a
safe place to prevent breakage.
4
Turn the manifold assembly
counterclockwise to remove it from the
rear bulkhead.
Note the positioning and placement of
each spray nozzle, for replacement
purposes.
5
Locate the clogged spray nozzle(s)
previously tagged and remove them from
the manifold by turning the spray nozzle
counterclockwise using a crescent
or deep socket wrench.
Turn the Spray Manifold
Counterclockwise
Turn Spray Nozzle
Counterclockwise
Spray Manifold