DC Puse Power Supply
【
DC-10-4P
】
8
APPENDIX
82
Command
Transmission data
Acknowledgment data
Command
Command name
Capacity
Description
Capacity Description
0x45
PARC
OSON
TIME
6byte
This is the command for setting PARC parameters.
The following parameters are set at once.
○
PARC_OSON_1ST_TIME
The parameter of CNFG/ARC/PARC/1ST in the menu.
The setting can be made in the unit of 0.1 us in the range
between current PULS_WIDTH to 20.0us.
The value in which the value intended to set is multiplied by
10 is transmitted.
20.0 →
200
:
0x00C8
○
PARC_OSON_2ND_TIME
The parameter of CNFG/ARC/PARC/2nd in the menu.
The setting can be made in the unit of 0.1 us in the range
between current PULS_WIDTH to 20.0us.
The value in which the value intended to set is multiplied by
10 is transmitted.
20.0 →
200
:
0x00C8
1byte
STATUS
○
PARC_OSON_ENABLE_TIME
The parameter of CNFG/ARC/PARC/ENBL in the menu.
It can be set in the range from 10 μs to 50,000 μs.
It can be set for every 10us.
50000
:
0xC350
The settings are made by transmitting 6-byte data in total by 2
bytes each.
Data is transmitted in the order of 1ST_TIME, 2ND_TIME,
and ENABLE_TIME.
* Be sure to send the values at once.
It is not possible to change specific parameters.
For any points without changes, send the values that are the
same as last time.
0x46
PULSE HARC
JDGE TIME
2byte
The judgment time for shifting into the HARC ARC can be set.
It can set in the range between 100 and 65,000us.
The minimum unit for setting is 10us.
Set the one-tenth the value to the set value.
100us
→
10
:
0x000A
65000us
→
6500
:
0x1964
1byte
STATUS
0x47
PULSE
LINK SET
2byte
Set the synchronization system No. for synchronized
operation using optical synchronization cables.
The parameter of PULS/LINK in the menu.
If the PULS_OSC (0x41) command is used to set one with
synchronization, be sure to set parameters.
Number of systems: 0 to 12 (0x00 to 0x0C)
System No.: 0 to 12 (0x00 to 0x0C)
Set the parameters as above.
However, set them within the range of the number of systems
≥ System No.
In addition, if the PULS_OSC(0x41) command is used to
select System with synchronization 1 (CMST), set the system
No. to 1. If System with synchronization 2 (CSLV) is selected,
set the system No. to 2.
Data is transmitted in the order of the number of systems and
the system No.
1byte
STATUS