Sensitivity Symptoms
146
DSQ II Hardware Manual
Thermo Scientific
No ions are visible in Tune after selecting “Restore Default Calibration”.
Possible Solutions
• Instrument is not calibrated. Run an automatic tune with all calibrations selected.
• No high
m/z
ion intensity because the RF is not tuned correctly. See
.
Poor high mass response.
Possible Solutions
• No high
m/z
ion intensity because the RF is not tuned correctly. See
• The ion volume, lenses, or prefilter are dirty. Clean the ion volume, lenses, and prefilter,
as described in
“Cleaning the Ion Source, Lenses and Prefilter”
.
• Poor high
m/z
ion ratios because the ion source temperature is too high. Reduce the ion
source temperature to reduce the amount of thermal decomposition and fragmentation of
your analyte.
• The conversion dynode is disconnected or faulty. Check the cable or replace the electron
multiplier/conversion dynode power supply.
• Reagent gas is leaking into the analyzer and suppressing the EI signal. Check for the
presence of CI reagent ions in the spectrum. If necessary, replace the CI reagent gas flow
module, as described in
“Replacing the CI Reagent Gas Flow Module”
.
Poor low mass sensitivity or
m/z
219 is base peak for calibration gas.
Possible Solutions
• Prefilter offset voltage is too negative. Adjust the prefilter offset closer to zero.
• The ion volume, lenses, or prefilter are dirty. Clean the ion volume, lenses, and prefilter,
as described in
“Cleaning the Ion Source, Lenses and Prefilter”
.
• Prefilter is misaligned or broken. Check the alignment of the prefilter.
• Lens voltages are incorrect. Select
Restore Default Tune
in
DSQ II Tune
.
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...