Maintaining the DSQ II
Cleaning the Ion Source Components
52
DSQ II Hardware Manual
Thermo Scientific
d. Attach the 3- and 8-pin connectors.
e.
Move the ion source left and right to ensure it is snapped in place. If it does not
move, it is in place. If the ion source is not in its groove properly, it could affect
performance and your ability to insert or remove the ion volume with an inlet valve.
f.
Remove the lens alignment tool.
g. Use the lens alignment tool to insert the ion volume. Make sure the arrow is facing up
and the ion volume is all the way in, which is typically when the flat surface of the ion
holder and heater board are exactly aligned. Any stainless steel that is outside the
board plane reduces ionization.
Cleaning the Prefilter
To clean the prefilter:
1. Remove the prefilter.
Figure 30.
Removing the Prefilter
1
Prefilter
4
Prefilter Clamp
2
Top Prefilter Connector
5
Bottom Prefilter Connector
3
Four Prefilter Clamp Screws
1
2
5
3
4
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...