High Vacuum Symptoms
Thermo Scientific
DSQ II Hardware Manual
137
Turbomolecular pump shuts off during operation.
Possible Solutions
• The fore pressure is too high. Check for leaks in the foreline.
• The cable connection between the turbomolecular pump and vacuum control board is
faulty. Check connections.
• The turbomolecular pump controller is faulty. Replace or repair as necessary.
• The turbomolecular pump power supply is faulty. Replace or repair as necessary.
• The turbomolecular pump is faulty. Replace the turbomolecular pump.
Unexpected system vent
occurs.
Possible Solutions
• A gross leak is present. Check for leaks.
• The system was vented through the inlet valve, there was a GC column break at the
transfer line or the foreline was cut. Shut off the DSQ II main circuit breaker. Allow the
system to vent and cool to room temperature.
Vacuum fault.
Possible Solutions
• A gross leak is present. Check for leaks in the transfer line, vent valve, high vacuum valve,
and rotary-vane pump connection.
• The ion gauge must be on (if you have one) and the pressure below 1x10
-3
Torr. Turn on
ion gauge from
DSQ II Tune
.
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...