2
Maintaining the DSQ II
Cleaning the Ion Source Components
44
DSQ II Hardware Manual
Thermo Scientific
5. Remove the ion volume from the ion source.
Figure 24.
Using the Ion Volume Tool to Remove the Ion Volume
a.
Hook the ion volume tool (PN 119270-0001) onto the ion
volume through the opening of the magnet yoke and white ion
source board.
b. Turn the ion volume tool 30° counter-clockwise to engage the ion
volume.
c.
Pull the ion volume straight back and through the opening.
d. Rotate the ion volume to disconnect it from the ion volume tool.
Figure 25.
Removing the Ion Volume from the Ion Volume Tool
6. Clean the ion volume, as described in "
Cleaning Stainless Steel Components" on page 34
.
If you are cleaning a CI ion volume, be sure to clean the small electron entrance hole with
a dental pick or old syringe needle. Aluminum oxide can get trapped in this hole, which
can adversely affect sensitivity.
1
Vacuum Manifold
4
Ion Volume Tool
2
Magnet Yoke
5
Ion Source Block (Ion Volume inside)
3
Ion Source Board
1
Ion Volume Tool
3
Ion Volume Arrow
2
Bayonet Pin Guide/Lock
4
Ion Volume
1
5
2
4
3
1
2
4
3
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...