2
Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
Thermo Scientific
DSQ II Hardware Manual
99
a.
Disconnect the five cables from the RF module.
b. Disconnect the three cables from the DC driver.
5. Remove the RF module.
a.
Remove the twelve screws holding the RF coil cover in place.
b. Remove the RF coil cover.
Figure 58.
RF Module with RF Coil Cover Removed
c.
Use forceps to disconnect the four RF feedthrough connectors.
d. Remove the bottom three and loosen the top three screws holding the RF module in
place.
e.
Remove the two screws holding the RF module to the front of the chassis.
f.
Lift the RF module up and out of its slots.
1
Four RF Feedthrough
Connectors
3
Two Screws Around the RF
Module
2
Six Screws Around the RF
Module
1
2
3
2
2