4
Vacuum System and Gas Inlets
Vacuum System
Thermo Scientific
DSQ II Hardware Manual
157
High Vacuum Pump (Split-Flow Turbomolecular Pump)
As an upgrade option, a differentially pumped configuration of DSQ II
is available. In a
differentially pumped instrument, the ion source is pumped separately from the quadrupole
and ion detector. Rather than using two individual turbomolecular pumps, a split-flow
turbomolecular pump is used. Such a pump uses less power, is more reliable, and less
expensive than two separate pumps.
With a differentially pumped system, you can use higher flow rates of gas (helium from the
GC and CI reagent gas) into the ion source without adversely affecting the performance of the
quadrupole. Sensitivity and resolution are optimum when the pressure in the quadrupole is
low.
Figure 77.
Split-flow Turbomolecular Pump
1
Electric Supply Connector (Not Shown)
2
Turbomolecular Pump Outlet
1
2
(behind pump)
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...