2
Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
80
DSQ II Hardware Manual
Thermo Scientific
Figure 43.
Removing the CI Reagent Gas Flow Module
a.
Disconnect the cable.
b. Disconnect the two gas fittings.
c.
Disconnect the foreline adapter tubing.
d. Remove the two screws holding the module in place.
6. Install the new CI reagent gas flow module.
a.
Insert the new module.
b. Reattach the screws.
c.
Reattach the foreline adapter tubing.
d. Reconnect the gas fittings.
e.
Reconnect the cable.
1
Gas Fittings
4
Cable (not shown)
2
Foreline Adapter Tubing (not
shown)
5
CI Reagent Gas Flow Module
(PN 23331-0091)
3
Two Screws
1
2
5
3
4
1