3
Troubleshooting
Communication Symptoms
Thermo Scientific
DSQ II Hardware Manual
125
3. Configure the TRACE GC Ultra handshaking parameter to allow the GC to become
ready to run a temperature program. Configure the TRACE GC Ultra as follows:
a.
Press <CONFIG> to display the Configure menu.
b. Press the down arrow key to Handshaking and press <ENTER> to display the Config
Handshaking menu.
c.
Press the down arrow key to Inhibit ready in and press <ENTER> to display the
Inhibit Ready menu.
d. Specify the Inhibit Ready parameter. The default setting is When High. This allows
normal operation of the DSQ II. If the GC method has not been downloaded from
Xcalibur
, the GC may not have received a signal from the DSQ II allowing it to
become ready. In this case, you can either send the GC method from
Xcalibur
or
select When low or Do not inhibit as the Inhibit Ready parameter.
Acquisition does not start as expected.
Possible Solutions
• Autosampler did not start. Verify autosampler methods and configuration. Verify that the
sample is present.
• Fore pressure is too high due to solvent peak. Reduce injection volume or extend
filament/multiplier delay time until after solvent peak. See
• TRACE GC Ultra did not start. Verify GC methods and configuration.Verify connection
between autosampler and GC.
• Heated zone setpoint is not attained. Refer to
.
• DSQ II Instrument Setup file has incorrect settings. Check start time in DSQ II
Instrument Setup file.
• DSQ II start mode is not properly configured. Verify that instrument is configured
properly. Refer to
Xcalibur
online help.
• Remote start cable to the GC is disconnected. Connect remote start cable.
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...