1
Basic Operations
Installing a Column into the GC/MS System
Thermo Scientific
DSQ II Hardware Manual
7
f.
Select
Pre/post column? yes (Y) or no (N)
.
• If a pre-/post-column is not present, select
N
.
• If a pre-/post-column is present, select
Y
. You must set the length and nominal
internal diameter of the pre-/post-column in the same valid ranges for the
column. The following two lines are added to the menu.
g. According to the physical characteristics of the column, the system calculates and
displays the relevant Column K-factor.
h. To evaluate the column, scroll to
Run column eval.?
and press
ENTER
.
The system will set the Column K factor by pressurizing the column to obtain a
nominal carrier flow of 5 mL/min. The display shows the pressure.
i.
Using the flowmeter, measure the carrier gas flow at the outlet of the column. Scroll
to
Measured Flow
and set the value of the flow.
j.
The following message will display if the operation was successful.
Pre/post column? Y
P/p col. L
10.00
P/p col. ID
0.530
LEFT
1
COLUMN INFO
Column K=
(0.8087)
Calc’d ID (0.0000)
Run Column eval?
1. These settings could also apply to a right
column.
EVALUATING L COLUMN
Pressure (162)
Measured Flow: 4.90
Use <STOP> to abort
Note
To abort column evaluation, you can press
STOP
.
L. COL. EVALUATION
COMPLETED
Calc’d ID 0.242
K. 0.9020
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...