Second, helium atoms (not ions) coming from the ion source cause noise if they reach the exit
lens or ion detector. This is known as neutral noise. Helium atoms cannot negotiate the bend
in the prefilter. Therefore, the noise is virtually eliminated. This leads to better signal-to-noise
ratio and lower limits of detection. In addition, higher GC helium flows can be used without
causing excessive amounts of noise that would blind an instrument without a curved prefilter
to low-level signals.
Figure 100.
Prefilter
Figure 101.
Quadrupole Mass Filter