Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
88
DSQ II Hardware Manual
Thermo Scientific
a.
Disconnect the Ethernet cable and J3 cable.
b. Remove the seven screws and two nuts holding the digital board cover in place.
c.
Remove the digital board cover.
d. Remove the six nuts holding the digital board in place.
e.
Carefully remove the digital board.
5. Install the new digital board (PN 119590-0010).
a.
Insert the new digital board.
b. Reattach the nuts.
c.
Reattach the digital board cover.
d. Reattach the screws.
e.
Reconnect the cables.
6. Reattach the left, top, and front covers of the DSQ II.
7. Restart the DSQ II.