Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
98
DSQ II Hardware Manual
Thermo Scientific
Replacing the Low Pass Filter Board
To replace the low pass filter, you will need:
• Forceps
• New Low Pass Filter board
• Screwdriver, Phillips #2
• Wrench, Allen, 2.5 mm
• Wrench, Allen, 3 mm
1. Cover the work area with a lint-free cloth.
2. Shut down and vent the DSQ II, as described in
"Shutting Down the GC/MS System"
.
3. Remove the left cover, as described in "
Removing the Left Cover" on page 29
4. Disconnect the RF module.
Figure 57.
Removing the RF Module
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug the DSQ II to
avoid an electrical shock.
1
RF Module
5
Three Cables to DC Driver
2
Two Cables to RF Detector Board
6
DC Driver
3
RF Coil Cover
7
Three Cables to Lens Interface and
RF Generator Board
4
12 Screws Around the RF Coil
Cover
3
7
5
6
4
2
4
1
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...