5
Ion Source and Inlet Valve
Inlet Valve (Upgrade)
180
DSQ II Hardware Manual
Thermo Scientific
I/R Tool and Guide Bar
The I/R (ion volume/removal) tool is used to insert and remove the ion volume and consists
of a shaft, handle, and guide ball. The I/R tool is designed to be used with the guide bar. The
guide ball located on the I/R tool fits inside a hole located on the guide bar, then slides
forward along the guide ball track. The guide bar is designed to accurately stop the I/R tool
the correct distance within the inlet valve in order to evacuate atmospheric pressure before
opening the ball valve. Once the ball valve is opened, the I/R tool is slid along the guide bar
track all the way into the vacuum manifold.
Figure 96.
I/R Tool and Guide Bar
1
Alignment Line
7
Guide Bar Track
2
Hold Position
8
Guide Bar
3
Release Position
9
Guide Bar First Stop
4
I/R Tool
10
Guide Ball Hole
5
Bayonet Lock
11
Guide Ball
6
Guide Bar Second Stop
12
Handle
4
3
2
5
6
7
9
10
8
11
12
1
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...