2
Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
Thermo Scientific
DSQ II Hardware Manual
81
7. Test the gas fittings for leaks.
a.
Turn on the CI reagent gas supply.
b. Use a leak detector to check for leaks on the high-pressure side of the CI reagent gas
flow module.
c.
Use the DSQ II to check for leaks on the low-pressure side of the CI reagent gas flow
module, as described in "
High Vacuum Symptoms" on page 135
.
d. Correct leaks by tightening or replacing the connections. Continue to check and
correct leaks until all connections are leak-free.
8. Reattach the left, top, and front covers of the DSQ II.
9. Restart the DSQ II.
10. Once the DSQ II has pumped down and is operational, check for vacuum leaks by
reading the convectron gauge and ion gauge and running an automatic tune.
Replacing the Conversion Dynode/Electron Multiplier Power Supply
To replace the conversion dynode/electron multiplier power supply, you will need a Phillips
#2 screwdriver.
1. Cover the work area with a lint-free cloth.
2. Shut down and vent the DSQ II, as described in
"Shutting Down the GC/MS System"
.
CAUTION - INSTRUMENT DAMAGE.
Do not use liquid soap leak detectors (such as
Snoop) to check for leaks. They may contaminate your system.
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug the DSQ II to
avoid an electrical shock.
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...