2
Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
Thermo Scientific
DSQ II Hardware Manual
89
Replacing the Electrometer Board
To replace the electrometer board, you will need:
• New Electrometer board (PN 119590-0500)
• Screwdriver, Phillips #1
• Wrench, Allen, 3 mm
• Wrench, open-ended, 17 mm
• Wrench, open-ended, 9/16-in.
1. Cover the work area with a lint-free cloth.
2. Shut down and vent the DSQ II, as described in
"Shutting Down the GC/MS System"
.
3. Remove the left cover, as described in "
Removing the Left Cover" on page 29
4. Remove the electrometer board.
Figure 49.
Removing the Electrometer Board
a.
Disconnect the cable.
b. Remove the calibration gas vial so you have space to remove the electrometer cover.
c.
Remove the two screws holding the electrometer board cover in place.
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug the DSQ II to
avoid an electrical shock.
1
Two Screws Around the
Electrometer Board Cover
4
Two Hex Bolts Around the Board
Cover
2
Cable
5
Cal Gas Vial
3
Electrometer Board Cover
1
5
4
2
3
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...