Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
72
DSQ II Hardware Manual
Thermo Scientific
4. Remove the analog board.
Figure 40.
Removing the Analog Board
1
Digital Board
8
P7 Connector on Electron Multiplier/
Conversion Dynode Power Supply
2
J1 Connector on RF Control Board
9
P8 Connector on RF Generator Board
3
P4/P5 Connector on DC Driver
10
P9 Connector on Power Module
4
J3 Connector on Low Pass Filter Board
11
P10 Connector on Vacuum Control
Board
5
J5 Connector on Analog Board
12
Nuts holding the Board in place
6
P2 Connector on Lens Interface Board
13
Analog Board (PN 119590-0022)
7
P6 Connector on CI Reagent Gas Flow
Module
4
10
13
12
6
3
1
12
5
8
2
11
7
12
9
12