5
Ion Source and Inlet Valve
Ion Source
174
DSQ II Hardware Manual
Thermo Scientific
Magnets and Magnet Yoke
Two magnets focus emitted electrons into a beam that spirals through the ion volume. The
spiraling of the beam ensure optimum ionization of the sample. The magnet yoke holds the
two magnets in the proper position above and below the ion source.
Position both magnets so the south pole is on top otherwise sensitivity will be poor. Electrons
are not focused into the ion volume if one magnet is upside down.
Figure 91.
Magnets and Magnet Yoke
1
Magnets
2
Magnet Yoke
1
2