Maintaining the DSQ II
Preparing the DSQ II for Maintenance
26
DSQ II Hardware Manual
Thermo Scientific
2. Shut down and vent the DSQ II, as described in
"Shutting Down the GC/MS System"
.
3. Remove the front cover.
4. Loosen the fastener located on the front of the DSQ II that attaches the top cover to the
chassis.
Figure 11.
Removing the Top Cover Fastener
5. Slide the top cover forward about 2.5 cm (1-inch).
6. Lift the top cover up and away from the DSQ II.
7. Reverse these steps to reinstall the top cover.
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug the DSQ II to
avoid an electrical shock
Top Cover Fastener