2
Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
114
DSQ II Hardware Manual
Thermo Scientific
Replacing the Turbomolecular Pump Power Supply
To replace the turbomolecular pump power supply, you will need a #2 Phillips screwdriver.
1. Cover the work area with a lint-free cloth.
2. Shut down and vent the DSQ II, as described in
"Shutting Down the GC/MS System"
.
3. Remove the right cover, as described in "
Removing the Right Cover" on page 27
.
4. Remove the turbomolecular pump power supply.
Figure 69.
Removing the Turbomolecular Pump Power Supply
a.
Disconnect the four cables (items 3, 4, 5, and 6 in ).
b. From the bottom of the DSQ II, remove the four screws holding the turbomolecular
power supply in place.
c.
Remove the turbomolecular pump power supply.
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug
the DSQ II to avoid an electrical shock.
1
Turbomolecular Pump Power
Supply (PN 76330-0230)
4
Connector to P2-2 Cable (+), red
wire
2
Four Screws Around the
Turbomolecular Pump Power
Supply
5
Connector to P1-2L Cable, brown
wire
3
Connector to P2-1 Cable (-), black
wire
6
Connector to P1-2N Cable, blue
wire
1
2
4
6
3
5
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...