2
Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
106
DSQ II Hardware Manual
Thermo Scientific
5. Remove the rear cooling fans from the fan bracket.
Figure 63.
Removing the Fans from the Bracket
a.
Disconnect the two fan cables.
b. Remove the four nuts holding the rear cooling fans in place.
c.
Remove the rear cooling fans.
6. Attach the new rear cooling fans to the fan bracket by reattaching the nuts. Make sure the
airflow indicator arrows on the new rear cooling fans point into the DSQ II (away from
the fan bracket) and the wires are pointing down.
7. Reconnect the fan cables.
8. Attach the two screws holding the fan bracket in place. Make sure the fan cables do not
get pinched or touch the fan blades.
9. Reattach the covers on the DSQ II.
10. Restart the DSQ II.
1
Fan Bracket
3
Four Nuts Around the Rear Cooling Fans
2
Two Rear Cooling Fans
4
Two Fan Cables
1
2
3
3
2
4
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...