2
Maintaining the DSQ II
Cleaning the Ion Source Components
56
DSQ II Hardware Manual
Thermo Scientific
8. Remove the components surrounding the ion source.
Figure 32.
Removing the Components Surrounding the Ion Source
9. Remove the source spacer, which contains the prefilter entrance lens.
10. Remove the lens stack.
11. Remove the magnet yoke, magnets, and thumbscrews as one unit and set them aside.
1
Source Spacer and Prefilter
Lens
6
Two Magnets
2
Lens Stack
7
Magnet Yoke
3
3-Pin Lens Connector
8
Two Thumbscrews
4
Filament
9
Ion Volume (Inside Ion Source)
5
8-Pin Lens Connector
10
Ion Source
2
3
4
5
7
8
1
9
10
6