2
Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
Thermo Scientific
DSQ II Hardware Manual
119
5. Install the new vent valve solenoid (PN 76461-1003). On the cable, the orientation of the
two plugs does not matter.
a.
Reattach the vent valve filter.
b. Attach the new vent valve solenoid.
c.
Reconnect the cable.
6. Reattach the left, top, and front covers of the DSQ II.
7. Restart the DSQ II.
8. Once the DSQ II has pumped down and is operational, check for vacuum leaks by
reading the convectron gauge and ion gauge pressures and running an automatic tune.