2
Maintaining the DSQ II
Preparing the DSQ II for Maintenance
Thermo Scientific
DSQ II Hardware Manual
31
Figure 16.
Removing the Screw Holding the Left Cover in Place
5. Slide the cover toward the rear of the instrument until the tabs are at the end of the
grooved slots and then pull the cover away.
6. Reverse these steps to reinstall the left cover.
RE
V
H IG H VO LTA G E
A ND H O T SU R FA C E S
WA R N IN G
C AU TIO N
Screw holding
left cover in
place
WARNING - INSTRUMENT DAMAGE.
The analog and digital boards are attached to
interior side of the left cover, so that cover is significantly heavier than the right cover. Do
not pull the cover too far away from the instrument when you open it or you could
damage the cables leading from the boards to other parts of the instrument.
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...