Maintaining the DSQ II
Preparing the DSQ II for Maintenance
Thermo Scientific
DSQ II Hardware Manual
27
Removing the Right Cover
To remove the right cover, you will need a Phillips #2 screwdriver and a 2.5 mm Allen wrench.
1. Cover the work area with a lint-free cloth.
Figure 12.
Components Accessed by Removing the Right Cover
2. Shut down and vent the DSQ II, as described in
"Shutting Down the GC/MS System"
.
3. Remove the front and top covers, as described in "
Removing the Front and Top Covers"
.
4. Remove the capillary column, as described in "
Preparing the DSQ II for Maintenance"
.
1
Inlet Valve
4
High Vacuum Pump
2
Lens Interface Board
5
Fore Pressure Gauge and Foreline Adapter
3
Transfer Line
6
Turbomolecular Pump Power Supply
2
3
4
5
1
6
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug the DSQ II to
avoid an electrical shock
Содержание DSQ II