Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
82
DSQ II Hardware Manual
Thermo Scientific
3. Remove the left cover, as described in "
Removing the Left Cover" on page 29
4. Remove the conversion dynode/electron multiplier power supply.
Figure 44.
Removing the Conversion Dynode/Electron Multiplier Power Supply
a.
Disconnect the ribbon cable.
b. Disconnect the electron multiplier cable and conversion dynode cable from the
bottom of the vacuum manifold.
c.
Remove the two large screws holding the power supply in place. There is
double-sided tape holding the back of the power supply to the chassis. Pull the power
supply out to break it loose from the tape.
5. Install the new conversion dynode/electron multiplier power supply (PN 119377-0001 or
119377-0002 with the PPINICI upgrade).
a.
Insert the power supply.
b. Reattach the screws.
c.
Reconnect the cables.
6. Reattach the left, top, and front covers of the DSQ II.
7. Restart the DSQ II.
1
Ribbon Cable
3
Screws Around the Power Supply
2
Electron Multiplier Cable
4
Conversion Dynode Cable
3
2
4
1
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...