2
Maintaining the DSQ II
Cleaning the Ion Source Components
46
DSQ II Hardware Manual
Thermo Scientific
Cleaning the Ion Source and Lenses
To clean the ion source and lenses:
1. Cover the work area with a lint-free cloth.
2. Shut down and vent the DSQ II, as described in
"Shutting Down the GC/MS System"
.
3. Remove the front and top covers, as described in "
Removing the Front and Top Covers"
.
4. Remove the vacuum manifold cover, as described in "
.
5. Disconnect the 3- and 8-pin connectors from the ion source. Do not disconnect them by
pulling on the wires.
6. Loosen the two thumbscrews on the magnet yoke.
7. Slide the source spacer slightly to the left, while pulling it away from the prefilter. Once it
is free, carefully slide it to the left so that it clears the end of the transfer line. Be careful
not to damage the capillary column extending out of the transfer line.
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug the DSQ II to
avoid an electrical shock.
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...