Troubleshooting
Using Diagnostics
122
DSQ II Hardware Manual
Thermo Scientific
Figure 73.
Diagnostics Flowchart
Running Tune Diagnostics
1. Open the Diagnostics window in
DSQ II Tune
.
a.
Open
Xcalibur
and click
Instrument Setup
.
b. Click the
DSQ II
icon to open the Method Editor.
c.
Click the
Tune
button.
d. Select
Diagnostics | Run Tests
to open the Diagnostics window.
2. Run diagnostics by clicking on the test you want to run.
Did the system fail while you
were running samples?
Did problems occur immediately
after you performed maintenance
on the instrument, data system,
or peripherals?
There may be a hardware
failure. Run diagnostics.
Did problems occur after you changed
the system's configuration, cables, or
peripherals?
The problem is probably mechanical,
not electrical. Verify that alignment,
configurations and cable connections
are correct before you run
Diagnostics
START
Run Diagnostics.
END
No
No
No
Yes
Yes
Yes
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...