Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
108
DSQ II Hardware Manual
Thermo Scientific
c.
Remove the RF detector board cover.
Figure 65.
Removing the RF Detector Board
d. Remove the five screws holding the RF detector board in place.
e.
Remove the RF detector board.
5. Install the new RF detector board (PN 119590-0270).
a.
Insert the RF detector board.
b. Reattach the screws that hold the RF detector board in place.
c.
Reattach the cover.
d. Reattach the screws that hold the cover in place.
e.
Reconnect the cables.
6. Reattach the left, top, and front covers of the DSQ II.
7. Restart the DSQ II.
1
RF Detector Board
(PN 119590-0270)
2
Five Screws Around the RF
Detector Board
1
2