3
Troubleshooting
RF and DC Control Symptoms
Thermo Scientific
DSQ II Hardware Manual
143
The Det RF and RF Mod lines are erratic.
Possible Solutions
• There is a loose cable. Check these cables:
• There is a problem with the RF control board, low pass filter board, the RF generator
board, or the RF coil. Contact Technical Support.
• There is high voltage arcing. Vent the system and check the quadrupole mass filter and
prefilter. Also, check the wires inside the RF coil box. Make sure the four wires going to
the RF feedthroughs are far from any surface and especially each other. Check for burn
marks or evidence of arcing. Clean components if necessary. Replace the components if
problems persist.
Rod DC diagnostic fails.
Possible Solutions
• If all tests fail, the DC driver is unplugged or faulty. Check all cables to the DC driver. If
necessary, replace the DC driver, as described in
• If A Rod DC and B Rod DC fail but Average DC passes, RF dip, RF gain, or RF null
may be adjusted incorrectly. Check RF dip and RF gain calibrations. The RF may not be
working properly or the DC driver is disconnected or faulty. Check the RF null (see the
DSQ II User’s Guide
).
Table 6.
Possible loose cables
Cables
Analog to RF Detector board
RF Detector to DC Driver and RF Generator board
Lens Interface to Low Pass Filter board
Analog to Low Pass Filter board
RF Generator to Low Pass Filter board
Cables from RF Coil to RF Detector Capacitor or RF feedthroughs
Cables from DC Driver to Low Pass Filter board
Analog to RF Generator board
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...