Maintaining the DSQ II
Maintaining the Rotary-Vane Pump
64
DSQ II Hardware Manual
Thermo Scientific
Adding Oil to a Rotary Vane Pump
To add oil to the rotary-vane pump:
1. Cover the work area with a lint-free cloth.
2. Shut down and vent the DSQ II, as described in
"Shutting Down the GC/MS System"
.
3. Add oil.
Figure 37.
Rotary Vane Pump Components
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug the DSQ II to
avoid an electrical shock.
Note
The pump oil level must be between the MIN and MAX marks on the oil-level
sight-glass for the pump to operate properly.
1
Foreline Vacuum Hose
8
Drain Plug
2
Inlet Port
9
MIN (minimum level mark)
3
Gas-Ballast Control
10
Mode Selector
4
Oil Filler Plugs
11
On/Off Switch
5
Outlet Port
12
Handle
6
MAX (maximum level mark)
13
Voltage Indicator
7
Oil-Level Sight-glass
14
Electrical Inlet Connector
14
3
1
2
5
6
10
9
11
13
12
7
4
8
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...