2
Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
Thermo Scientific
DSQ II Hardware Manual
117
a.
Disconnect all the cables from the Vacuum Control board.
Figure 71.
Removing the Vacuum Control Board Mounting Screws
b. Remove the four mounting screws holding the vacuum control board in place.
c.
Remove the vacuum control board.
5. Install the new vacuum control board (PN 119590-0031).
a.
Insert the new vacuum control board.
b. Reattach the screws that hold the vacuum control board in place.
c.
Reconnect all the cable to the vacuum control board.
6. Reattach the left, top, and front covers of the DSQ II.
7. Restart the DSQ II.
1
Four Vacuum Control Board Mounting Screws
1