Maintaining the DSQ II
Cleaning the Ion Source Components
Thermo Scientific
DSQ II Hardware Manual
49
Figure 28.
Disassembling the Lens Stack
a.
Remove the lens clip by pinching the ends with your fingers or forceps.
b. Remove the lenses and lens spacers from the ceramic lens holder.
13. Clean the exterior and interior surface of the lenses, as described in
. Pay particular attention to the areas inside the tube and
around the hole of each lens.
14. Clean ONLY the ceramics with Aqua Regia, hydrogen flame, or muffle furnace at 600 °C.
1
Lens Clip
4
Lens 2
2
Lens 1
5
Lens 3
3
Lens Spacer
6
Lens Holder
1
2
3
4
3
5
6
Disassembled
Assembled