2
Maintaining the DSQ II
Maintaining the Rotary-Vane Pump
Thermo Scientific
DSQ II Hardware Manual
69
7. Add fresh oil.
a.
Add oil to the oil filler plug opening halfway between the MIN and MAX level
marks.
b. If the oil level goes above the MAX level mark, remove the drain plug and drain the
excess oil from the pump.
8. Reassemble the rotary-vane pump.
a.
Reattach the oil filler plug.
b. Place the rotary-vane pump on the floor.
c.
Reconnect the foreline vacuum hose.
9. Plug the rotary-vane pump into the mechanical pump outlet on the back of the DSQ II.
10. Restart the DSQ II.
WARNING - FIRE HAZARD.
Use only factory-approved rotary-vane pump oil. Using
unapproved rotary-vane pump oil may result in a fire or explosion.