Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
Thermo Scientific
DSQ II Hardware Manual
85
Replacing the DC Driver Fuse
To replace the DC driver fuse, you will need a 2.5 mm Allen wrench.
1. Cover the work area with a lint-free cloth.
2. Shut down and vent the DSQ II, as described in
"Shutting Down the GC/MS System"
.
3. Remove the left cover, as described in "
Removing the Left Cover" on page 29
4. Remove the DC driver assembly cover.
Figure 46.
Removing the DC Driver
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug the DSQ II to
avoid an electrical shock.
1
Screws Around the DC Driver
Cover
3
DC Driver
2
DC Driver Cover
2
3
1