Maintaining the DSQ II
Preparing the DSQ II for Maintenance
32
DSQ II Hardware Manual
Thermo Scientific
Removing the Vacuum Manifold Cover
If you need to clean the ion source, prefilter, quadrupole, or ion detector, you will need to
remove the vacuum manifold cover first.
To remove the vacuum manifold cover, you will need:
• Gas, clean and dry
• Gloves, clean, lint- and powder-free
• Lint-free cloth
• Screwdriver, Phillips #2
• Wrench, Allen, 2.5 mm
1. Cover the work area with a lint-free cloth.
Figure 17.
Components Accessed By Removing the Vacuum Manifold Cover
2. Shut down and vent the DSQ II, as described in
"Preparing the DSQ II for Maintenance"
.
CAUTION - INSTRUMENT DAMAGE.
Keep the vacuum manifold cover attached to the
instrument while performing maintenance. Even small amounts of dust on the electron
multiplier, conversion dynode, or quadrupole can cause excessive noise in your mass
spectra.
1
Electron Multiplier
4
Prefilter
2
O-ring
5
Ion Source
3
Quadrupole
6
Transfer Line
5
3
2
6
1
4
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug the DSQ II to
avoid an electrical shock
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...