1
Basic Operations
Installing a Column into the GC/MS System
Thermo Scientific
DSQ II Hardware Manual
5
a.
Unwind the column about a half turn for the inlet and two turns for the DSQ II.
b. Wipe about 100 mm (4 in.) of the column with a tissue soaked in methanol.
c.
Insert the column through the injector nut and ferrule (open end up). Wipe the
column again with a tissue soaked in methanol.
d. Use a scoring wafer to score and break the column about 2.5 cm (1 in.) from the end.
Use a magnifying glass to check for an even, flat cut. Repeat if necessary.
e.
Insert the column into the injector so that the end of the column is the proper
distance (splitless = 64 mm, split = 40 mm, PTV = 30 mm) from the back of the
injector nut.
f.
Verify the position with the column measuring tool.
g. Use your fingers to tighten the injector nut and give it an additional quarter turn
with the wrench.
h. Use a scoring wafer to score and break the column outlet about 2.5 cm (1 in.) from
the end.
i.
Power on the TRACE GC Ultra.
2. Set up the TRACE GC Ultra.
a.
Set the oven and injector temperatures to 30 °C.
b. Set the injector flow to 1.0 mL/min.
c.
Turn off vacuum compensation (under the Right or Left Carrier menu).
d. Dip the column outlet in a small vial of methanol. Bubbles indicate there is flow
through the column.
e.
Allow the column to purge for at least 10 minutes.
3. Perform column characterization.
a.
Raise the oven and injector temperatures to 50 °C and allow them to stabilize. Then
confirm that the carrier gas is on.
Note
Wear clean, lint- and powder-free gloves when you handle the column and
injector ferrule.
Tip
Slide a septum on the column before the injector nut to make it easier to
measure the proper distance between the nut and end of the column.
CAUTION - INSTRUMENT DAMAGE:
Do not allow the clear plastic component to
exceed 80 °C. Otherwise, it will melt and damage the instrument.
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...