2
Maintaining the DSQ II
Cleaning the Ion Source Components
Thermo Scientific
DSQ II Hardware Manual
41
b. Turn the I/R tool handle counter-clockwise toward the hold position, as indicated by
the
icon.
c.
Withdraw the tool until the guide ball fits in the groove furthest from the instrument.
d. Push the inlet valve lever down to close it.
e.
Loosen the inlet valve knob.
f.
Remove the I/R tool with the ion volume attached to it.
6. Clean the ion volume.
a.
Press the ion volume forward into the tip of the I/R tool and rotate it to disconnect
the bayonet pins from the pin guides.
Figure 23.
Disconnecting the Ion Volume from the I/R Tool
I
WARNING - ELECTRICAL SHOCK HAZARD.
The ion volume may be very hot, so let it
cool to room temperature before handling it.
1
I/R Tool
6
Spring Washer
2
Ion Volume
7
Ion Volume Alignment Arrow
3
I/R Tool Alignment Arrow
8
Bayonet Pins
4
Pin Guide
9
Bayonet Lock
5
Ion Volume Holder
1
2
3
4
5
6
7
9
8
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...