Contamination Symptoms
Thermo Scientific
DSQ II Hardware Manual
129
Chemical background due to rhenium oxide (
m/z
185/187, 201/203, 217/219, 233/235,
250/252).
Possible Solutions
These series of rhenium oxide ions come from oxidation of the rhenium filament wire due to
the introduction of air into the ion source while the filament is on. Check for air leaks and
remedy. See
Spectra are observed due to solvents.
Possible Solutions
• There is residual solvent from a cleaning procedure or laboratory background
contamination. When you finish performing a cleaning procedure, allow cleaned
components to dry thoroughly. Warm parts in the GC oven to dry off residual solvent.
• The observed compounds have been introduced through sample injection. Ultimate
source is either a sample solvent or the autosampler rinsing solvent. Optimize GC
method to separate solvent peak from area of interest in the chromatogram.
Table 3.
Solvent Spectra
Solvents
Acetone (
m/z
43, 58, 59)
Hexane (
m/z
41, 43, 56, 57, 58, 85, 86)
Methanol (
m/z
31, 32)
Methylene chloride (
m/z
84/83)
Toluene (
m/z
91, 92)
Trichloroethane (
m/z
151, 153)
Xylene (
m/z
105, 106)
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...