Troubleshooting
Filament and Lens Control Symptoms
Thermo Scientific
DSQ II Hardware Manual
131
Emission current is unstable.
Possible Solutions
• The filament is near the end of its life span. Replace the filament, as described in
“Replacing the Ion Source Filament”
• There is a fault in the ion source board. Replace the ion source board, as described in
“Cleaning the Ion Source, Lenses and Prefilter”
.
• There is a problem with the analog board. If necessary, replace the analog board, as
described in
Short filament lifetime.
Possible Solutions
• Change the emission current setting to:
−
CEI – 25 μA.
−
PCI – 50 μA.
−
NCI – 50 μA.
• There is an air leak contributing to short filament lifetime. Check for leaks; repair if any
are found. See
.
• The filament is on during solvent peak. Increase filament/multiplier delay time until
solvent peak has passed.
• High emission current is being used. Use lower emission current to extend lifetime.
Diagnostics indicate flat lens response.
Possible Solutions
• The power supply to the lens has a fault. Check regulator board in the power module;
replace if necessary, see
• There is a fault with the lens drivers. Replace the analog board, as described in
.
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...