Maintaining the DSQ II
Cleaning the Ion Source Components
Thermo Scientific
DSQ II Hardware Manual
43
9. Restore the inlet valve.
a.
Reattach the inlet valve plug. Keep the EI or CI label pointing upward to indicate
which type of ion volume is installed in this instrument. If there is an EI/CI ion
volume installed in this instrument, position the label so that EI and CI are left to
right.
b. Remove the guide bar by rotating it 90° counter-clockwise and sliding it out of the
guide bar opening.
10. Close the Insert/Remove Probe window.
Cleaning the Ion Volume Without an Inlet Valve
To clean the ion volume without an inlet valve, you will need:
• Cleaning supplies for stainless steel parts
• Gloves, clean, lint- and powder-free
• Ion volume tool
• Lint-free cloth
1. Cover the work area with a lint-free cloth.
2. Shut down and vent the DSQ II, as described in
"Shutting Down the GC/MS System"
.
3. Remove the front and top covers, as described in "
Removing the Front and Top Covers"
.
4. Remove the vacuum manifold cover, as described in "
.
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug the DSQ II to
avoid an electrical shock.
Note
Wear clean, lint- and powder-free gloves when handling any vacuum manifold
components.
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...