2
Maintaining the DSQ II
Preparing the DSQ II for Maintenance
Thermo Scientific
DSQ II Hardware Manual
35
2. Rinse the components with clean water. Use a clean applicator or toothbrush to remove
the aluminum oxide slurry. Do not let the slurry dry on the stainless steel because it is
difficult to remove.
3. Sonicate the components in warm detergent.
a.
Use forceps to place the components in a beaker of warm detergent.
b. Place the beaker and its contents in an ultrasonic bath for five minutes.
c.
Rinse the components with tap water to remove the detergent.
4. Sonicate the components in deionized water.
a.
Use forceps to place the components in a beaker of deionized water.
b. Place the beaker and its contents in an ultrasonic bath for five minutes.
c.
If the water is cloudy after sonicating, pour out the water, add fresh water, and place
the beaker and its contents in a ultrasonic bath again for five minutes. Repeat until
the water is clear.
5. Sonicate the components in acetone.
a.
Use forceps to place the components in a beaker of acetone.
b. Place the beaker and its contents in an ultrasonic bath for five minutes.
c.
Use forceps to transfer the components to a beaker of fresh acetone.
d. Place the beaker and its contents in an ultrasonic bath for five minutes.
6. Blow clean, dry gas on the components to remove the acetone.
7. Use forceps to place the components in a beaker covered with aluminum foil. Dry the
components in a 100
o
C oven for 30 minutes. Allow the components to cool before
reassembling them.
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...