5
Ion Source and Inlet Valve
Inlet Valve (Upgrade)
Thermo Scientific
DSQ II Hardware Manual
175
Inlet Valve (Upgrade)
The inlet valve mounts on the front of the vacuum manifold and is the vacuum interlock
mechanism for the DSQ II. The inlet valve is a vacuum sealed valve through which you insert
and remove the ion volume assembly using an I/R (insertion/removal) tool.
With the use of the I/R tool and guide bar, the inlet valve lets the ion volume be removed or
installed in the ion source without having to vent the vacuum manifold to atmospheric
pressure. Once excess pressure is removed by opening the inlet valve solenoid, the inlet valve
can be opened that allows the I/R tool to be inserted into the vacuum manifold. Excess
pressure introduced into the inlet valve when inserting the I/R tool is removed through the
foreline hose connection back to either a primary or secondary rotary-vane pump.
The inlet valve consists of the following components:
• Ball valve
• Inlet valve lever
• Guide bar housing
• I/R tool and guide bar
• Inlet valve plug
• Inlet valve seal
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...