2. What Is what
ABL800 FLEX Operator's Manual
Analyzer status
Analyzer status
at a glance
The working condition of the analyzer is continuously monitored during its
operation.
To evaluate the analyzer status at a glance before a measurement, use the following
facilities:
•
parameter bar
•
color of the traffic light on the
Analyzer Status
button.
Analyzer Status button
Red = cannot be used; Yellow = unreliable; Green = OK
To enter the
Analyzer Status
screen, press
Menu
-
Analyzer Status.
Analyzer Status
elements
Traffic light color of the
Analyzer Status
button is determined by the traffic light
colors of the following status elements:
Status element
The colors indicate…
Calibration Status
G
REEN
– OK.
Y
ELLOW
– error(s) in the last calibration and/or cal
schedule reminders.
RED –
error-prone parameters are repressed.