5. Quality control
ABL800 FLEX Operator's Manual
Quality control result,
Continued
High Crea
Check
(for the ABL8x7
FLEX analyzers
only)
The high Crea check result of the ABL8x7 FLEX analyzers can be retrieved as
follows:
Press
Result
on
the
Quality Control Identification
screen
from
the
Menu
>
Analyzer status
>
Quality Control
from the Quality Control log
The initial
c
Crea concentration in the Cleaning Met II Solution is approximately
1500
μ
mol/L when the container is installed in the analyzer. The exact
concentration is entered from the barcode when the solution is installed in the
analyzer (the value shown as
c
Crea on the
High cCrea Check
screen above) .
With time, creatinine in the solution transforms into creatine, the speed of this
transformation depends on the ambient temperature.
Expected
c
Crea concentration in the Cleaning Met II Solution
The upper/lower drift limits for
c
Crea is
±
225
μ
mol/L, i.e. if the
initial cCrea is 1500
μ
mol/L, the corresponding control ranges will
bwe 1275-1725
μ
mol/L.
Expected
c
Crea concentration in the Cleaning Met II Solution at
the time of the High Crea check has been performed
Measured
c
Crea concentration in the Cleaning Met II Solution at
the time of the High Crea check has been performed
1500
μ
mol/L
c
Crea
Crea
dif
Time
Continued on next page
5-12
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...