ABL800 FLEX Operator's Manual
3. Installation and setup
Calibration setup,
Continued
Calibration
schedule
This program allows you to schedule automatic calibrations, manual tHb
calibration and automatic cleaning (the first automatic cleaning also includes
quality control of the Crea A and B electrodes for high creatinine values on the
ABL8x7 FLEX analyzers), and select a calibration to be performed after a sample
measurement.
To edit the schedule for automatic calibrations/cleaning and tHb calibration, do the
following:
Step Action
1.
Highlight the desired calibration or cleaning on the screen.
2.
Press the
Edit
button.
24-hour scale
shows the time for
each scheduled
automatic
calibration or
cleaning.
List of activities:
calibrations and
cleaning.
Activity after
each
measurement
Use the arrow buttons to select start time for calibrations and cleaning
and the intervals between each activity.
Continued on next page
3-33
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...