ABL800 FLEX Operator's Manual
3. Installation and setup
Quality control setup,
Continued
NOTICE:
Once the Fixed SD is activated, you cannot update the control ranges to
the limits that are narrower than those determined by the Fixed SD, for both single-
parameter and multiple-parameter updates.
QC ranges
(continued)
This program allows you to select the following for the
Quality Control
Identification
screen during a measurement:
QC input setup
•
mandatory temperature entry by the operator
•
the default temperature always displayed (unless altered by the operator).
To select the settings, do the following:
Step Action
1.
Activate the
Mandatory temperature
check button.
A
"
will appear next to an empty temperature box on the
Quality
Control Identification
screen during each quality control
measurement and a temperature must be entered, otherwise the result
cannot be retrieved.
2.
Highlight the default temperature in the "Default temperature" box
and enter a default temperature on the keypad; confirm with
Enter
.
The value in
°
C or
°
F is automatically entered on the
Quality Control
Identification
screen during measurement. The temperature can be
altered for a particular measurement but will return to the default
setting for future measurements.
Continued on next page
3-39
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...